| Pump Specifications
|
| Mechanism |
Peristaltic |
| IV set type |
General |
| Infusion Modes |
Rate mode |
| Flow rate range |
0.1~1500ml/h |
| VTBI |
0.1~999.9 ml, the increment is 0.1 ml;
1000-9999ml, increment is 1ml |
| Accumulated volume |
0.1-9999ml |
| Increments |
0.1ml/h(0.1-999.9ml/h)
1ml/h(1000-1500ml/h) |
| Accuracy (± %) |
≤±5% |
| KVO |
0.5 ml/h |
| Bolus rate |
0.10-1500ml/h |
| Purge rate |
600ml/h |
| Occlusion level |
3 levels |
| Occlusion pressure range |
Low 150±125 mmHg (20kPa);
Medium 525±125 mmHg (70kPa);
High 900±180 mmHg (120 kPa) |
| Air bubble level |
5 levels |
| Air bubble size |
Single bubble detection: 50, 100, 250, 500, 800l μl
Accumulated bubble detection: 0.1-4.0ml/h, adjustable |
| Display
|
| Screen |
2.5” LCD monochrome screen, 132 x 32 pixels |
| Display info. |
Flow rate, current infusion,VTBI, total volume, IV set brand, pressure limit, battery capacity, drugs, remaining time, alarms |
| Functions
|
| The change flow rate during infusion |
● |
| Front-panel lockout |
● |
| Standby |
● |
| Pause |
● |
| Silence |
● |
| Drip rate function |
● (optional) |
| Alarm
|
| Real-time alarm display |
● |
| Auto self-test system |
● |
| Infusion finish |
● |
| Infusion near end |
● |
| IV set disengaged |
● |
| Occlusion |
● |
| Air in line |
● |
| Flow error |
● |
| Door open |
● |
| System malfunction |
● |
| System abnormal |
● |
| Low battery |
● |
| Battery empty |
● |
| Reminder |
● |
| Standby end |
● |
| No AC Power |
● |
| Interface
|
| Data port |
● RS232 |
| Nurse call jack |
● |
| DC input |
● |
| Battery
|
| Battery type |
Lithium |
| Battery operating time |
>4 hours@25ml/h |
| Battery recharge time |
≤6 hours@100% |